Air Flow Transducer MODEL 6333

Air Flow Transducer MODEL 6333

Best remote censing device simultaneously measuring air velocity, temperature, humidity for IoT monitoring

CHARACTERISTIC


6333 is an user-friendly stationary-type anemometer which is easy to built in other devices such as DIN rail. Both analog and digital output are available. Particularly, analog output part has 3 systems which are compatible to air velocity, temperature, humidity, and simultaneous output of the three parameters is available. Moreover, 6333 can be installed in a variety of places such as building, residence, factory, data center, etc, and it enables you to remotely take measurement data.

  • 15 types of compatible probes for various applications
  • The smallest size miniature probe in the industry available
  • Measurement range from Min. 0.01m/s to Max. 50.0m/s with high responsivity and accuracy
  • High heat resistance (Temperature range: 0~120℃) enables to accurately measure even in severe conditions
  • Compatible with both analog output and digital output
  • Current output and digital output are switchable

USE


  • For remote monitoring air velocity, temperature, humidity in various kinds of applications

SPECIFICATIONS


Specification of Main Unit
Product Name Airflow Transducer
Model Number 6333-00
Air Velocity Range 0.01~25.0m/s (Probe type: 0975-00, 0975-21, 0975-31)
0.01~30.0m/s (Probe type: 0976-03, 0976-04, 0976-05, 0976-07, 0976-13, 0976-14, 0976-15, 0976-17)
0.01~50.0m/s (Probe type: 0972-00, 0973-00, 0975-09, 0975-10)
Temperature Range 0~60℃ (Probe type: 0975-31)
0~100℃ (Probe type: 0975-21, 0976-15)
0~120℃ (Probe type: 0976-05)
Humidity Range 5~95% (Probe type: 0975-31)
Resolution Air Velocity: 0.01m/s (*0.01~9.99m/s), Air Velocity: 0.1m/s (*10.0~50.0m/s), Temperature: 0.1℃, Humidity: 0.1%RH
Display Unit m/s, ℃, %RH
Analog Output Current output and voltage output are switchable by DIP switch (*Standard setting is current output)
Current output: DC4~20mA (Max. load: 250Ω)
Voltage output: DC0~5V
Interface RS485
Power Supply DC12~24V
Please select either one of 3 options below
1. Connection with the other unit
2. AC adaptor (90W) or AC adaptor (160W) (*if two or more hub units are added)
3. DC power cable
Power Supply Please select either one from the following 2 options
1. AC adaptor: AC100~240V (*input), DC12V (*output)
2. External power supply to terminal block: DC12~24V
Power Consumption Approx. 3W (*if power supply voltage: 12V, Air velocity: approx. 10m/s, probe type: 0976-03)
Operational Environment Temperature: 0~60℃ (*no condensation), Humidity: 2~95%RH
Storage Environment Temperature: -20~70℃ (*no condensation), Humidity: 2~95%RH
Size Approx. W100 x D60 x H25 (mm)
Weight Approx. 200g
Standard Accessories Instruction manual, terminal block
Warranty For 1 year after the purchase
Model Parameter Directivity Velocity Temperature Humidity Operational
Temperature
Temperature
Compensation
Meas. Range Accuracy Responsivity Meas. Range Accuracy Meas. Range Accuracy
0972-00
0973-00
Velocity Directional 0.01~50.0m/s ±2% of reading or ±0.02m/s,
whichever is greater
1 sec N/A N/A N/A N/A 0~100℃ 0~100℃
±5% of reading
0975-00 Velocity Omi-directional 0.01~25.0m/s ±2% of reading or ±0.02m/s,
whichever is greater
3 sec N/A N/A N/A N/A 0~100℃ 0~100℃
±5% of reading
0975-09
0975-10
Velocity Omi-directional 0.01~50.0m/s ±2% of reading or ±0.02m/s,
whichever is greater
3 sec N/A N/A N/A N/A 0~100℃ 0~100℃
±5% of reading
0976-03
0976-04
0976-07
Velocity Omi-directional 0.01~30.0m/s ±3% of reading or ±0.02m/s,
whichever is greater
3 sec N/A N/A N/A N/A 0~120℃ 0~120℃
±5% of reading
0976-13
0976-14
0976-17
Velocity Omi-directional 0.01~30.0m/s ±2% of reading or ±0.02m/s,
whichever is greater
3 sec N/A N/A N/A N/A 0~100℃ 0~100℃
±5% of reading
0976-05 Velocity,
Temperature
Omi-directional 0.01~30.0m/s ±3% of reading or ±0.02m/s,
whichever is greater
3 sec 0~120℃ ±0.5% N/A N/A 0~120℃ 0~120℃
±5% of reading
0976-15 Velocity,
Temperature
Omi-directional 0.01~30.0m/s ±2% of reading or ±0.02m/s,
whichever is greater
3 sec 0~100℃ ±0.5% N/A N/A 0~100℃ 0~100℃
±5% of reading
0975-21 Velocity,
Temperature
Omi-directional 0.01~25.0m/s ±2% of reading or ±0.02m/s,
whichever is greater
3 sec 0~100℃ ±0.5% N/A N/A 0~100℃ 0~100℃
±5% of reading
0975-31 Velocity,
Temperature,
Humidity
Omi-directional 0.01~25.0m/s ±2% of reading or ±0.02m/s,
whichever is greater
3 sec 0~60℃ ±0.5% 5~95%RH 5~80%RH: 3%RH
80~90%RH: 5%RH
0~60℃ 0~100℃
±5% of reading
  • MODEL 0972-00
    0972-00
  • MODEL 0973-00
    0973-00
  • MODEL 0975-00
    0975-00
  • MODEL 0975-09
    0975-09
  • MODEL 0975-10
    0975-10
  • MODEL 0975-31
    0975-31
  • MODEL 0976-03
    0976-03
  • MODEL 0976-04
    0976-04
  • MODEL 0976-05
    0976-05

ACCESSORIES


  • Instruction manual
  • Terminal block

OPTIONAL ITEMS


  • Probe
  • Probe cable (0.5m, 2m, 5m, 10m, 20m, 40m)
  • AC adaptor
  • DIN rail fixing base
  • Probe fixing jig
  • Double-sided tape for probe (model 0976)

MOVIE