Best remote censing device simultaneously measuring air velocity, temperature, humidity for IoT monitoring
6333 is an user-friendly stationary-type anemometer which is easy to built in other devices such as DIN rail. Both analog and digital output are available. Particularly, analog output part has 3 systems which are compatible to air velocity, temperature, humidity, and simultaneous output of the three parameters is available. Moreover, 6333 can be installed in a variety of places such as building, residence, factory, data center, etc, and it enables you to remotely take measurement data.
Specification of Main Unit | |
Product Name | Airflow Transducer |
---|---|
Model Number | 6333-00 |
Air Velocity Range | 0.01~25.0m/s (Probe type: 0975-00, 0975-21, 0975-31) |
0.01~30.0m/s (Probe type: 0976-03, 0976-04, 0976-05, 0976-07, 0976-13, 0976-14, 0976-15, 0976-17) | |
0.01~50.0m/s (Probe type: 0972-00, 0973-00, 0975-09, 0975-10) | |
Temperature Range | 0~60℃ (Probe type: 0975-31) |
0~100℃ (Probe type: 0975-21, 0976-15) | |
0~120℃ (Probe type: 0976-05) | |
Humidity Range | 5~95% (Probe type: 0975-31) |
Resolution | Air Velocity: 0.01m/s (*0.01~9.99m/s), Air Velocity: 0.1m/s (*10.0~50.0m/s), Temperature: 0.1℃, Humidity: 0.1%RH |
Display Unit | m/s, ℃, %RH |
Analog Output | Current output and voltage output are switchable by DIP switch (*Standard setting is current output) Current output: DC4~20mA (Max. load: 250Ω) Voltage output: DC0~5V |
Interface | RS485 |
Power Supply | DC12~24V Please select either one of 3 options below 1. Connection with the other unit 2. AC adaptor (90W) or AC adaptor (160W) (*if two or more hub units are added) 3. DC power cable |
Power Supply | Please select either one from the following 2 options 1. AC adaptor: AC100~240V (*input), DC12V (*output) 2. External power supply to terminal block: DC12~24V |
Power Consumption | Approx. 3W (*if power supply voltage: 12V, Air velocity: approx. 10m/s, probe type: 0976-03) |
Operational Environment | Temperature: 0~60℃ (*no condensation), Humidity: 2~95%RH |
Storage Environment | Temperature: -20~70℃ (*no condensation), Humidity: 2~95%RH |
Size | Approx. W100 x D60 x H25 (mm) |
Weight | Approx. 200g |
Standard Accessories | Instruction manual, terminal block |
Warranty | For 1 year after the purchase |
Model | Parameter | Directivity | Velocity | Temperature | Humidity | Operational Temperature |
Temperature Compensation |
||||
---|---|---|---|---|---|---|---|---|---|---|---|
Meas. Range | Accuracy | Responsivity | Meas. Range | Accuracy | Meas. Range | Accuracy | |||||
0972-00 0973-00 |
Velocity | Directional | 0.01~50.0m/s | ±2% of reading or ±0.02m/s, whichever is greater |
1 sec | N/A | N/A | N/A | N/A | 0~100℃ | 0~100℃ ±5% of reading |
0975-00 | Velocity | Omi-directional | 0.01~25.0m/s | ±2% of reading or ±0.02m/s, whichever is greater |
3 sec | N/A | N/A | N/A | N/A | 0~100℃ | 0~100℃ ±5% of reading |
0975-09 0975-10 |
Velocity | Omi-directional | 0.01~50.0m/s | ±2% of reading or ±0.02m/s, whichever is greater |
3 sec | N/A | N/A | N/A | N/A | 0~100℃ | 0~100℃ ±5% of reading |
0976-03 0976-04 0976-07 |
Velocity | Omi-directional | 0.01~30.0m/s | ±3% of reading or ±0.02m/s, whichever is greater |
3 sec | N/A | N/A | N/A | N/A | 0~120℃ | 0~120℃ ±5% of reading |
0976-13 0976-14 0976-17 |
Velocity | Omi-directional | 0.01~30.0m/s | ±2% of reading or ±0.02m/s, whichever is greater |
3 sec | N/A | N/A | N/A | N/A | 0~100℃ | 0~100℃ ±5% of reading |
0976-05 | Velocity, Temperature |
Omi-directional | 0.01~30.0m/s | ±3% of reading or ±0.02m/s, whichever is greater |
3 sec | 0~120℃ | ±0.5% | N/A | N/A | 0~120℃ | 0~120℃ ±5% of reading |
0976-15 | Velocity, Temperature |
Omi-directional | 0.01~30.0m/s | ±2% of reading or ±0.02m/s, whichever is greater |
3 sec | 0~100℃ | ±0.5% | N/A | N/A | 0~100℃ | 0~100℃ ±5% of reading |
0975-21 | Velocity, Temperature |
Omi-directional | 0.01~25.0m/s | ±2% of reading or ±0.02m/s, whichever is greater |
3 sec | 0~100℃ | ±0.5% | N/A | N/A | 0~100℃ | 0~100℃ ±5% of reading |
0975-31 | Velocity, Temperature, Humidity |
Omi-directional | 0.01~25.0m/s | ±2% of reading or ±0.02m/s, whichever is greater |
3 sec | 0~60℃ | ±0.5% | 5~95%RH | 5~80%RH: 3%RH 80~90%RH: 5%RH |
0~60℃ | 0~100℃ ±5% of reading |