The smallest 6-channel portable particle counters in the industry with 28.3L/min (3905), 50L/min (3910), 100L/min (3920) sampling flow rate.
Counting efficiency50±20% at 0.3μm50±20% at 0.5μmConcentration limit500,000 particles/ft³ at 10% coincidence loss300,000 particles/ft³ at 10% coincidence lossSampling time6sec ~ 23hours 59min 59sec (configurable)Interval time6sec ~ 23hours 59min 59sec (configurable)InterfaceEthernet, USBPowerLi-ion battery or AC adapter (AC100 to 240V)BatteryRemovable/rechargeable Li-ion batteries (2 battery Slots)Operating environmentTemperature: 10 ~ 40℃ / Humidity: 20 ~ 85%RH (*no condensation)Storage environmentTemperature: -10 ~ 50℃ / Humidity: up to 98%RH (*no condensation)
Portable Particle Counters | |||
---|---|---|---|
Model | 3905 | 3910 | 3920 |
Light source | Laser diode | ||
Particle size | 0.3, 0.5, 1.0, 3.0, 5.0, 10.0 (μm) | 0.5, 0.7, 1.0, 3.0, 5.0, 10.0 (μm) | |
Flow Rate | 28.3 L/min | 50 L/min | 100 L/min |
Zero count level | Less than 1 count per 5 minutes | ||
Printer | Built-in thermal printer | ||
Display | 6.4 inch color touch screen | ||
Enclosure | Stainless Steel | ||
Data storage | Up to 10,000 data | ||
Security | Password authentication and management, secure data, and audit trails (meets 21 CFR Part 11 compliance) | ||
Reports | ISO146441-1, GB/T16292-2010, FEDERAL STANDARD, BRITISH STANDARD, EU GMP | ||
Standards | ISO 21501-4, JIS B9921, CE | ||
Dimensions | Approx. 200 × 205 × 200 (mm) (*without handle) | Approx. 350 × 210 × 210 (mm) (*without handle) | |
Weight | Approx. 4.5kg (*without battery) | Approx. 5.0kg (*without battery) | |
Warranty | 2 years after the purchase |
We have several varieties including: handheld particle counters in 3 and 6 channel versions; 6-channel portable particle counters with up to a 50LPM flow rate; a highly specialized condensation particle counter that features a 0.015 micron sensitivity making it ideal for Nano-sized particle counting; and a fixed cleanroom monitoring system that can be integrated with a PC for 24/7 monitoring.
Annual calibration is recommended to ensure accuracy. However, this can vary based on how often the particle counter is used and the environments it’s used it. Also, some companies have an SOP in place that specifies how often the instrument will need to be serviced. More specifically, regulatory compliances may dictate the calibration interval.
You can measure clean air between -4 and 158°F (-20 to 70°C) in an environment with no visible condensation. (Please note: the temperature range given is for the probes)
Windows XP (32-bit only), Vista, & Windows 7.
Absolutely. Just make sure you select a counter that can measure the specific sized particles you need to measure. Handheld model 3888 and portable models 3905 & 3910 are pre-programmed with helpful standards modes that simplify the certification process.
Our portable units are 21 CFR part 11 compliant, and an IQ/OQ is available for purchase if needed.
No, you don’t need a separate instrument. Model numbers 3889, 3905 & 3910 can measure these parameters just by adding optional probes to the particle counter.